Tuesday 3 October 2017

TASK 1 FOR LABVIEW CHAPTER 1

Hi Assalammualaikum today our lecturer has given to us a task regarding chapter 1 labView
So, here we gooooooooooo

The task is about

1. Use DAQ USB 6210 to simulate the input the input signal generate from an analogue trainer
2. connect the DAQ to your PC's and ensure your PC's communicate with the DAQ
3. Use an Analogue input (A1) channel, ranging from A1 (0-15) as an input to your analogue trainer      output
4. Generate an output from the analogue trainer to simulate an output display of the following :

I. Triangle with 3 Vp-p
II. Square with 1.5 Vamp
III. Sinusoidal with 2.55 Vrms 

-----------------------------------------------------------------------------------------------------------------------

First of all, setup all the things needed, DAQ module and also the LabView software.
Connect the DAQ module with computer or laptop. Then, connect the ground and channel input from the DAQ module to the function generation (ground to ground & channel input to output function generator ). 




  • open the LabView >> click New >> New VI 
  • at the block diagram >> right click >> and choose Express >> Input >> DAQ Assistant >> drag to the block diagram 
  • wait for it to build :)
  • then it will pop up new DAQ Assistant window like picture below  


regarding the picture, choose Acquire Signal >> Analogue Input >> Voltage 





Next, choose the input channel that we need to use 





check the connection trough connection diagram 




  • then, set 1)Acquisition Mode to Continuous, 2)Sample to Read and Rate(Hz) with '1 to 1 ratio' 3)Terminal Configuration set to RSE. 
  • After set it all then click OK. A few seconds later will pop up new windows and click YES So wait it to install.


* So before proceed we have to add the waveform graph. At the front panel >> Right Click >> choose Graph >> choose Waveform Graph  (picture shown below)*

Next, focused back to the task I. 
  • At the function generator, set to the 3Vp-p (adjust the amplitude knob) which means 1.5 v at upper side and 1.5 at the bottom side, and the type of wave knob move to the triangle. 



task II 
  • adjust the knob type of wave to square 



task III

  • next change the knob to the sinosaidal wave 
  • and set the vrms with value of  2.55v. By using formula to find vrms Vp = (Vrms)/ √2

then we get Vp = 3.6 v 


The End...

So for you guys information, we are group of three which consist of me, Aina and Syamiha 
so check it out for the next task either in this link nursyamiha.blogspot.com or this link ainasyafinifattah.blogspot.com .

see yaa :)) 








Sunday 17 September 2017

CHAPTER 3: MODIFYING A VI CREATED FROM A TEMPLATE


Complete the following steps to create a VI that generates, analyzes, and
displays a signal.

1.  In the Getting Started window, click New to display the New dialog
box.
2.  From the Create New list, select VI»From Template»Tutorial
(Getting Started)»Generate, Analyze, and Display. This template
VI simulates a signal and analyzes it for its root mean square (RMS)

3.  Click the OK button or double-click the name of the template to create
a VI from the template.
4.  If the Context Help window is not visible, press the <Ctrl-H> keys to
display the window. (Mac OS X) Press the <Command-Shift-H> keys.
5. Display the block diagram by pressing the <Ctrl-E> keys.
6. Move the cursor over the Amplitude and Level Measurements Express
VI, shown at left.
The Context Help window displays information about the behavior of
the Express VI.
Keep the Context Help window open. It will provide useful
information as you complete the rest of this exercise.
7. Display the front panel and remove the RMS indicator, shown at left.
You will not use the RMS functionality of the Amplitude and Level
Measurements Express VI for this exercise. However, you can use the
Generate, Analyze, and Display template VI with the RMS
functionality in the future to reduce development time.
8. Display the block diagram and remove any broken wires that result
from removing the RMS indicator. To remove all broken wires from
the block diagram, you can press the <Ctrl-B> keys.
9. Then return to the front panel window and right-click the waveform
graph indicator. Select Properties from the shortcut menu. The Graph
Properties dialog box appears.
10. On the Appearance page, place a checkmark in the Visible checkbox
in the Label section and enter Unfiltered Signal in the text box.
11. Click the OK button to save the configuration and close the Graph
Properties dialog box.
12. Run the VI.
The signal appears in the graph.
13. Click the STOP button to stop the VI.





ADDING A SIGNAL 

The Simulate Signal Express VI simulates a sine wave by default. You can
customize the simulated signal by changing the options in the Configure
Simulate Signal dialog box.

Complete the following steps to create an additional simulated signal that
adds uniform white noise to the sine wave.

1. On the block diagram, use the Positioning tool to select the Simulate
Signal Express VI.
Hold down the <Ctrl> key and click and drag to create an additional
Simulate Signal Express VI on the block diagram. (Mac OS X) Hold
down the <Option> key and drag. (Linux) You also can hold down the
middle mouse button and drag.
2. Release the mouse button to place the copied Simulate Signal Express
VI below the original Simulate Signal Express VI. LabVIEW updates
the name of the copied Simulate Signal Express VI to Simulate
Signal2.
3. Double-click the Simulate Signal2 Express VI to display the
Configure Simulate Signal dialog box.


4. Select Sine from the Signal type pull-down menu.
5. Enter 60 in the Frequency (Hz) text box.
6. Enter 0.1 in the Amplitude text box.
7. Place a checkmark in the Add noise checkbox to add noise to the sine
signal.
8. Select Uniform White Noise from the Noise type pull-down menu.
9. Enter 0.1 in the Noise amplitude text box.
10. Enter -1 in the Seed number text box.
11. In the Timing section, select the Run as fast as possible option.
12. In the Signal Name section, remove the checkmark from the Use
signal type name checkbox.
13. Enter 60 Hz and Noise in the Signal name text box.




When you change the signal name in the Configure Simulate Signal
dialog box, LabVIEW changes the name of the signal output on the
block diagram. Changing the signal name makes it easier for you to
identify the signal type when you view the Express VI on the block
diagram.
14. Click the OK button to save the current configuration and close
the Configure Simulate Signal dialog box



ADDING TWO SIGNAL

To add two signals together to create one signal, you can use the Formula Express VI. Rather than merging two signals on one graph, the Formula Express VI adds both signals together to create a single signal on the graph. You can use this Express VI to add noise to a signal.


Complete the following steps to add the 60 Hz and Noise signal to the Sine signal.

1. In the block diagram window, triple-click the wire that connects the
Sine output of the Simulate Signal Express VI to the Signals input of
the Amplitude and Level Measurements Express VI and to the
Unfiltered Signal indicator. Remove the wire.
2. On the Functions palette, click the Search button to search for the
Formula Express VI, shown at left, and add it to the block diagram
between the Simulate Signal Express VIs and the Amplitude and Level
Measurements Express VI. The Configure Formula dialog box
appears.




Note LabVIEW displays a folder glyph to the left of subpalettes in the search results and
displays a light blue glyph to the left of Express VIs in the search results.
3. In the Label column, change the label for X1 to Sine and the label for
X2 to 60 Hz and Noise.
The Formula Express VI automatically enters the first input, Sine, in
the Formula text box.
4. Click the + button and then the X2 button to add Sine and 60 Hz and
Noise together in the Formula text box.





5. Click the OK button to save the current configuration and close the
Configure Formula dialog box.
6. Use the Wiring tool to wire the Sine output of the Simulate Signal
Express VI to the Sine input of the Formula Express VI.
7. Wire the 60 Hz and Noise output of the Simulate Signal 2 Express VI
to the 60 Hz and Noise input of the Formula Express VI.
8. Wire the Result output of the Formula Express VI to the Unfiltered
Signal indicator and to the Signals input of the Amplitude and Level
Measurements Express VI.
9. Display the front panel by pressing the <Ctrl-E> keys.
10. Run the VI.


The signal with added noise appears in the graph.

11. Click the STOP button to stop the VI.
12. Select File>>Save As and save the VI as Analysis.vi in an easily
accessible location.



FILTERING SIGNAL 


You can use the Filter Express VI to process signals through filters and
windows.
Complete the following steps to configure the Filter Express VI to filter the
signal using an infinite impulse response (IIR) filter.
1. Display the block diagram window and remove the wire that connects
the Result output of the Formula Express VI to the Signals input of the
Amplitude and Level Measurements Express VI.
2. Remove all broken wires that result from removing the wire.




3. Search for the Filter Express VI, shown at left, and add it to the block
diagram between the Simulate Signal2 Express VI and the Amplitude
and Level Measurements Express VI. The Configure Filter dialog
box appears.



4. In the Filter Specifications section, change the Cutoff Frequency (Hz) to 25.


5. Click the OK button to save the configuration and close the Configure
Filter dialog box.
6. Display the front panel.
7. Click the Unfiltered Signal waveform graph indicator and press the
<Ctrl> key while you drag with the Positioning tool to create an
additional waveform graph indicator.



8. Triple-click the Unfiltered Signal 2 label above the new waveform
graph indicator and enter Filtered Signal to change the label of
the indicator. You also can change the label on the Appearance page
of the Graph Properties dialog box.



9. On the block diagram, wire the Result output of the Formula Express
VI to the Signal input of the Filter Express VI and to the Unfiltered
Signal waveform graph indicator.
10. Wire the Filtered Signal output of the Filter Express VI to the Signals
input of the Amplitude and Level Measurements Express VI aand to the
input of the Filtered Signal waveform graph indicator.


11. Select File >> Save. The block diagram of the Analysis VI should appear similar to Figure 3-3.



MODIFYING THE APPEARANCE OF GRAPH 

You can use the Display Format page of the Graph Properties dialog box to specify how the scales of the x-axis and y-axis appear on the graph. Complete the following steps to change the format of the x-axis and y-axis of the Unfiltered Signal and Filtered Signal graphs.

1. In the front panel window, right-click the Unfiltered Signal graph indicator and select Properties from the shortcut menu. The Graph Properties dialog box appears.
2. On the Display Format page, select Time (X-Axis) from the top pull-down menu.
3. Select the Default editing mode option.
4. In the Type list, select Automatic formatting.
5. In the Digits field, enter 6 and select Significant digits from the Precision Type pull-down menu.
6. Place a checkmark in the Hide trailing zeros checkbox.
7. Select Amplitude (Y-Axis) from the top pull-down menu and repeat steps 3–6 so the y-axis configuration matches the x-axis configuration.

8. On the Scales page, select Amplitude (Y-Axis).

9. Remove the checkmark from the Autoscale checkbox.
10. Enter -2.5 in the Minimum text box and 2.5 in the Maximum text
box.

11. Click the OK button to save the configuration and close the Graph
Properties dialog box.
12. Repeat steps 1–11 to configure the Filtered Signal graph indicator.
The x-axis and the y-axis on the Unfiltered Signal and Filtered
Signal graph indicators change to reflect the new configuration




ANALYZING THE AMPLITUDE OF A SIGNAL 

You can use the Amplitude and Level Measurements Express VI to analyze the voltage characteristics of a signal. Complete the following steps to reconfigure the Express VI to measure the
peak-to-peak amplitude values of the signal.

1. On the block diagram, double-click the Amplitude and Level
Measurements Express VI to display the Configure Amplitude and
Level Measurements dialog box.
2. In the Amplitude Measurements section, remove the checkmark
from the RMS checkbox.
3. Place a checkmark in the Peak to peak checkbox. Peak to peak
appears in the Results section with the corresponding value of the
measurement.



4. Click the OK button to save the current configuration and close
the Configure Amplitude and Level Measurements dialog box.
The RMS output of the Amplitude and Level Measurements Express
VI changes to reflect the new Peak to Peak output, shown at left.
You will use the Peak to Peak output in a later exercise.


CONTROLLING THE SPEED OF EXECUTION 

To plot the points on the waveform graphs more slowly, you can add a time
delay to the block diagram. A time delay slows the speed at which a VI
runs.
Complete the following steps to control the speed at which the VI runs.
1. On the block diagram, search for the Time Delay Express VI.
2. Place the Time Delay Express VI inside the While Loop. The
Configure Time Delay dialog box appears.


3. Enter 1.000 in the Time delay (seconds) text box and click the OK

Button
4. Display the front panel and run the VI. The VI runs more slowly.
The loop iterates once every second.
5. Stop the VI.


Another way to control the speed of the VI is to alter the rate of data acquisition. On the block diagram, double click the Simulate Signal Express VI to display the Configure Simulate Signal dialog box. Locate the Timing section in the dialog box. The Timing section contains a number of ways to alter the rate of data acquisition and the speed at which a VI runs. For example, one of the default settings of the VI is Simulate Acquisition Timing. This means that the VI mimics the acquisition rate of a hardware device. You can select Run as fast as possible to display data more quickly. In the Samples per second (Hz) text box, the default value is 1000, while the default value in the Number of Samples text box is 100. This means that the VI will output 100 data
points spanning 0.1 second. You can change these values to change the amount of data the VI displays, as well as the rate at which the VI displays the data.


ADDING A WARNING LIGHT 

If you want a visual cue to indicate when a value exceeds a specified limit,
you can use a warning light.
Complete the following steps to add a warning light to the VI.
1. Display the Controls palette by right-clicking any blank space in the
front panel window.
2. On the Express palette, select the LEDs palette.
3. Select the round LED indicator and add it to the front panel to the left
of the waveform graphs.
 4. Double-click the Boolean label above the LED and enter Warning to
change the label of the LED.


You will use this LED in a later exercise to indicate when a value has
exceeded its limit.
5. Select File»Save As to display the Save As dialog box.
6. Read the various dialog box options. Select the Copy and Substitute
copy for original options to create a copy of the original VI and
immediately edit the copy.
7. Click the Continue button and save the VI as Warning Light.vi in
an easily accessible location.



SETTING A WARNING LEVEL LIMIT


To specify the value at which you want the LED to light, use the Comparison Express VI.
Complete the following steps to compare the peak-to-peak value to a limit you set.
1. On the block diagram, search for the Comparison Express VI and place it to the right of the Amplitude and Level Measurements Express VI. The Configure Comparison dialog box appears.




2. In the Compare Condition section, select the > Greater option.

3. In the Comparison Inputs section, select Value and enter 2 in the Value numeric control to assign a constant value at which you want the LED to light.
4. Click the OK button to save the current configuration and close the Configure Comparison dialog box. The name of the Comparison Express VI changes to reflect the operation of the Express VI, shown at left. Greater indicates that the Express VI does a greater than comparison.
5. Wire the Peak to Peak output of the Amplitude and Level Measurements Express VI to the Operand 1 input of the Greater Express VI.
6. Move the cursor over the wire that connects the Peak to Peak output to the Operand 1 input.
7. When the Positioning tool appears, right-click the wire that connects
the Peak to Peak output to the Operand 1 input and select Create»
Numeric Indicator from the shortcut menu.


A Peak to Peak terminal, shown at left, appears on the block diagram. If the Peak to Peak terminal appears to be on top of the wires between the Express VIs, move the Express VIs and Peak to Peak terminal around to create more space. For example, move the Peak to Peak terminal into blank space above the Express VIs.



WARNING THE USER 

After you specify the values at which you want the LED to light, you must
wire the LED to the Greater Express VI.

Complete the following steps to provide a visual cue when the peak-to-peak
value of the signal exceeds a specified limit.
1. In the block diagram window, move the Warning terminal to the right of the Greater Express VI. Make sure the Warning terminal is inside the While Loop, as shown in Figure 3-4.
2. Wire the Result output of the Greater Express VI to the Warning terminal.






3. Display the front panel. A numeric indicator labeled Peak to Peak appears in the front panel
window. This indicator displays the peak-to-peak value of the signal.
4. Run the VI. When the peak-to-peak value exceeds 2.0, the Warning indicator lights.
5. Click the STOP button to stop the VI.
6. Save the VI.


CONFIGURING A VI TO SAVE DATA TO A FILE 


To store information about the data a VI generates, use the Write To Measurement File Express VI.
Complete the following steps to build a VI that saves peak-to-peak values and other information to a LabVIEW data file.

1. Search for the Write To Measurement File Express VI and add it to the block diagram below and to the right of the Amplitude and Level Measurements Express VI.

The Configure Write To Measurement File dialog box appears. The Filename text box displays the full path to the output file, test.lvm. A .lvm file is a tab-delimited text measurement file you
can open with a spreadsheet application or a text-editing application. LabVIEW saves data with up to six digits of precision in a .lvm file. LabVIEW saves the .lvm file in the default LabVIEW Data
directory. LabVIEW installs the LabVIEW Data directory in the default file directory of the operating system. When you want to view the data, use the file path displayed in the Filename text box to access the test.lvm file.



2. In the Configure Write to Measurement File dialog box, locate the If a file already exists section and select the Append to file option to write all the data to the test.lvm file without erasing any existing data in the file.
3. In the Segment Headers section, select the One header only option
to create only one header in the file to which LabVIEW writes the data.
4. Enter the following text in the File Description text box: Sample of peak to peak values. LabVIEW appends the text you enter in this text box to the header of the file.
5. Click the OK button to save the current configuration and close the Configure Write To Measurement File dialog box.



SAVING THE DATA TO A FILE 

When you run the VI, LabVIEW saves the data to the test.lvm file. Complete the following steps to generate the test.lvm file.

1. On the block diagram, wire the Peak to Peak output of the Amplitude and Level Measurements Express VI to the Signals input of the Write To Measurement File Express VI.
2. Select File»Save As and save the VI as Save Data.vi in an easily accessible location.
3. Display the front panel and run the VI.
4. Click the front panel STOP button.
5. To view the data you saved, open the test.lvm file in the LabVIEW Data directory with a spreadsheet or text-editing application. The file has a header that contains information about the Express VI.
6. Close the file after you finish looking at it and return to the Save Data VI.






ADDING A BUTTON THAT STORES DATA WHEN CLICKED 


If you want to store only certain data points, you can configure the Write To Measurement File Express VI to save peak-to-peak values only when a user clicks a button. Complete the following steps to add a button to the VI and configure how the button responds when a user clicks it.
1. Display the front panel and search the Controls palette for a rocker button. Select one of the rocker buttons and place it to the right of the waveform graphs.


2. Right-click the rocker button and select Properties from the shortcut menu to display the Boolean Properties dialog box.
3. Change the label of the button to Write to File.
4. On the Operation page of the Boolean Properties dialog box, select Latch when pressed from the Button behavior list. Use the Operation page to specify how a button behaves when a user
clicks it. To see how the button reacts to a click, click the button in the Preview Selected Behavior section.
5. Click the OK button to save the current configuration and close the Boolean Properties dialog box.
6. Save the VI.


SAVING DATA WHEN PROMPTED BY A USER 

Complete the following steps to build a VI that logs data to a file when the user clicks a button on the front panel.

1. In the block diagram window, double-click the Write To Measurement File Express VI to display the Configure Write To Measurement File dialog box.
2. Change the filename test.lvm in the Filename text box to Selected Samples.lvm to save the data to a different file.
3. Click the OK button to save the current configuration and close the Configure Write To Measurement File dialog box.
4. Right-click the Signals input of the Write To Measurement File Express VI and select Insert Input/Output from the shortcut menu to
insert the Comment input.


5. Right-click the Comment input of the Write To Measurement File Express VI and select Select Input/Output»Enable from the shortcut menu to replace the Comment input with the Enable input.



The inputs and outputs of an Express VI appear in a predetermined order when you add new inputs and outputs. To select a specific input, you might need to add an input first, and then change the input to the specific one you want to use by right-clicking the input and selecting Select Input/Output from the shortcut menu.

6. Move the Write to File terminal to the left of the Write To
Measurement File Express VI.
7. Wire the Write to File terminal to the Enable input of the Write To
Measurement File Express VI.



VIEWING SAVED DATA 

Complete the following steps to view the data that you save to the Selected Samples.lvm file.
1. Display the front panel and run the VI. Click the Write to File button
several times.
2. Click the STOP button.
3. Open the Selected Samples.lvm file with a spreadsheet or text-editing application.
The Selected Samples.lvm file differs from the test.lvm file. test.lvm recorded all the data generated by the Save Data VI, whereas Selected Samples.lvm recorded data only when you clicked the Write to File button.


4. Close the file after you finish looking at it.

5. Save and close the VI.























TASK 1 FOR LABVIEW CHAPTER 1

Hi Assalammualaikum today our lecturer has given to us a task regarding chapter 1 labView So, here we gooooooooooo The task is about 1....